Notified Body Under the MDR: How to Find and Select One

Notified body selection under the EU MDR using NANDO scope codes

The MDR has applied since 26 May 2021, with stricter designation requirements for notified bodies than under the directives. Implementing Regulation (EU) 2017/2185 of 23 November 2017 sets the codes that define what each notified body may assess, and MDCG 2019-6 Rev.5 (February 2025) sets out what a notified body may and may not do before an application. On 16 December 2025, the Commission proposed changes to re-certification and audits; that is a proposal, not law.

A notified body whose scope does not fit your device costs time. This article explains when you need a notified body, how to check scope codes in NANDO, what happens before and after you apply, and how certificates and surveillance work.

What a notified body is

According to the European Commission, a notified body is an organisation designated by a Member State, or by another country under a specific agreement, to assess conformity where EU legislation requires a third party. It must be impartial and independent, and must keep what it receives confidential. Manufacturers may choose any legally designated notified body, not only one in their own country.

The Commission lists designated bodies in NANDO, with their identification numbers and the tasks they are designated for. The Commission's notified bodies page links to the MDR list.

When you need one

Device Notified body involvement
Class I, not sterile, no measuring function, not a reusable surgical instrument Not required
Class I sterile, with a measuring function, or a reusable surgical instrument Required, limited to those aspects
Class IIa, IIb and III Required

Sources: MDR Article 52; Commission MDR manufacturer factsheet.

Our articles on classification of medical devices and the CE mark for a medical device cover the class and the conformity assessment route. IVD notified bodies are designated under the IVDR in the same way; our article on IVD CE marking covers that route.

Checking scope codes in NANDO

A notified body can only certify devices inside its designated scope. The scope is expressed in codes from Implementing Regulation 2017/2185, explained in MDCG 2019-14:

Code type Reflects Per device
MDA / MDN Design and intended purpose Exactly one, the highest applicable in the list
MDS Specific characteristics None or several
MDT Technologies or processes One or several

Source: MDCG 2019-14.

Standalone software is coded MDA 0315. According to MDCG 2019-14, the manufacturer proposes the codes and the notified body makes the final assignment during the application review. A designation can also carry conditions, so read the full NANDO record, not only the code list.

In our view, check scope codes before price. A notified body without the right code cannot take your device, however good its reputation.

Before you apply

According to MDCG 2019-6 Rev.5 (sections I.6.2 and I.6.3):

  • a notified body does not carry out a gap analysis or a partial assessment of your quality management system before an application;

  • it cannot act as your consultant;

  • before quoting, it checks the regulatory status, classification and proposed codes;

  • a structured dialogue covers the contract, the documents to submit, languages, timelines and fees, at no extra fee.

The contract follows the application review (MDCG 2019-14). Standard fees must be made public (MDR Article 50). MDCG 2022-14 (August 2022) adds measures such as hybrid audits and schemes for SMEs and first-time applicants. If you withdraw an application, that information is shared through EUDAMED (Article 53(2)).

In our view, use the structured dialogue to clarify what needs to be fulfilled, not how. Arrive with the intended purpose, the classification rationale, a proposed code list and the state of the quality management system. Our article on SaMD technical documentation covers what the file itself contains.

After certification

Element Rule Source
Certificate validity Maximum five years, then re-certification SWD(2025) 1050; MDCG 2019-13
Surveillance audits Annual SWD(2025) 1050
Unannounced audit At least one per five-year cycle SWD(2025) 1050
Sampling before certification, class IIa At least one representative device per category MDCG 2019-13
Sampling before certification, class IIb At least one representative device per generic device group MDCG 2019-13
Technical documentation during validity At least one assessed each year; whole range covered over the validity period MDCG 2019-13

Sources: SWD(2025) 1050; MDCG 2019-13 Rev.1.

According to MDCG 2019-6 Rev.5, a supplier's ISO 13485 certificate alone is not enough to show supplier control. Changing notified body is possible and voluntary (MDR Article 58; MDCG 2019-6 Rev.5, section IV.4).

The December 2025 proposal (COM(2025) 1023) would replace re-certification with periodic review. Until it is adopted, the five-year cycle applies.

A hypothetical example

Hypothetical example: a Dutch SaMD company has decision-support software classified as class IIa under Rule 11.

  1. Shortlist: it filters the NANDO MDR list for notified bodies holding MDA 0315 and reads each record, including any conditions.

  2. Dialogue: it holds a structured dialogue with its shortlist on classification, codes and fees.

  3. Application: it applies to one notified body and signs the contract after the application review.

  4. Certification: at least one representative device per category is assessed before the quality management system certificate is issued.

  5. Afterwards: it plans for annual surveillance, technical documentation sampling, an unannounced audit and re-certification within five years.

Conclusion

A notified body is required for class IIa, IIb and III devices, and for some class I devices. The certificate is only valid inside the body's designated scope, so the scope check comes first.

In our view, confirm the codes in NANDO, use the structured dialogue for clear requirements, and plan for the whole certificate cycle from the start.

If you need help preparing for a notified body, our regulatory strategy team can help, or contact MedQAIR.

Frequently asked questions

Can a notified body help me prepare my submission?

No. According to MDCG 2019-6 Rev.5, it cannot carry out a gap analysis or act as your consultant. The structured dialogue covers the contract, documents, languages, timelines and fees.

How do I check whether a notified body covers standalone software?

Look for MDA 0315 in its NANDO record under the MDR, and read any conditions attached to the designation.

How long is an MDR certificate valid?

A maximum of five years, followed by re-certification.

What is an unannounced audit?

An audit without prior notice. At least one takes place in each five-year certificate cycle.

Can I change notified body?

Yes. A voluntary change is possible under MDR Article 58, with practical arrangements described in MDCG 2019-6 Rev.5.

Latest Blogs

November 16, 2026

The MDR has applied since 26 May 2021. In the US, the Quality Management System Regulation (QMSR) took effect on

November 15, 2026

The MDR has applied since 26 May 2021, with stricter designation requirements for notified bodies than under the directives. Implementing

November 14, 2026

Since 1 October 2023, all 510(k) submissions, unless exempted, must be submitted electronically using eSTAR (FDA). For FDA's fiscal year

logo

Unlock Your Quick Guide to AI Act
Compliance!

Explore AI-enabled SaMD requirements with our easy step-by-step guide.

Cookies help us improve your experience on our website. By using our site, you consent to the use of cookies as described in this policy.