Since 2 February 2026, FDA's Quality Management System Regulation (21 CFR Part 820) has incorporated ISO 13485:2016 by reference. Even so, FDA states that it "will not require certificates of conformance to ISO 13485 and will not issue" them (QMSR FAQ, Q13). ISO itself "does not perform certification or issue certificates".
A certificate is evidence from a third party, not a licence from a regulator. This article covers who certifies, how accreditation works, the audit cycle, and what a certificate does and does not do under the EU MDR and the US QMSR.
ISO does not certify
According to ISO's certification page, "ISO does not perform certification or issue certificates", and "a company or organization cannot be certified by ISO". ISO develops the standard; an independent certification body audits against it.
On its ISO 13485 page, ISO states that certification "is not a requirement of the standard", but that third-party certification "can demonstrate to regulators" that the requirements are met. Our article on the ISO 13485 quality management system covers the requirements themselves.
Certification and accreditation
ISO defines the two terms differently:
Certification: "The provision by an independent body of written assurance (a certificate) that the product, service or system in question meets specific requirements."
Accreditation: "The formal recognition by an independent body, generally known as an accreditation body, that a certification body operates according to international standards."
| Body | Role | Source |
|---|---|---|
| ISO | Publishes ISO 13485 and the CASCO standards used by certification bodies; does not certify | ISO certification page |
| Certification body | Audits the organisation against ISO 13485 and issues the certificate | ISO certification page |
| Accreditation body | Accredits certification bodies; accreditation is not compulsory | ISO certification page |
| Global ACI | Has assumed the former roles of IAF and ILAC | ISO certification page |
Source: ISO, Certification.
Because accreditation is not compulsory, a certificate alone does not show that the issuer is accredited. ISO points to the national accreditation body or IAF CertSearch to find accredited bodies, and to IAF CertSearch or the certification body to verify a certificate.
Choosing a certification body
ISO's checklist:
evaluate several certification bodies;
check that the body follows the relevant CASCO standard;
check whether it is accredited.
The ISO logo may not be used to show certification.
In our view, choose the certification body for the markets you need, not the price. Its scope and accreditation decide where the certificate counts.
Audit stages and the three-year cycle
The audit structure below is from the MDSAP Audit Approach (MDSAP AU P0002.009), which refers to ISO/IEC 17021-1:2015. It describes the MDSAP approach, not requirements of ISO 13485 itself.
| Audit | Purpose | When | Source |
|---|---|---|---|
| Stage 1 | Documentation review and preparedness for Stage 2 | Initial audit | P0002.009; ISO/IEC 17021-1:2015 Cl. 9.3.1.2 |
| Stage 2 | Implementation and effectiveness, at all sites to be on the certificate | Initial audit, after Stage 1 | P0002.009; ISO/IEC 17021-1:2015 Cl. 9.3.1.3 |
| Surveillance | Ongoing assessment | Each of the two years after certification | P0002.009 |
| Recertification | Renewed assessment | Third year | P0002.009 |
| Special or unannounced | Additional assessment | Possible during the cycle | P0002.009 |
Source: MDSAP AU P0002.009.
P0002.009 describes a "three (3) year audit cycle". FDA's FAQ (Q13) says third-party audits "occur annually based on a three-year plan". According to the MDSAP Audit Approach page, certificates are issued after an initial or re-certification audit, and "not normally" after surveillance.
In our view, agree the full audit programme before signing, not just the initial audit.
United States
According to FDA's QMSR FAQ (Q13):
FDA will not require or issue ISO 13485 certificates;
a certificate "will not exempt a manufacturer from an FDA inspection";
FDA inspections do not follow the MDSAP audit approach.
In our view, a certificate shows the QMS was audited against the standard; it is not proof of QMSR compliance. Our article on 21 CFR Part 820 explains what the QMSR requires.
European Union
MDCG 2019-6 Rev.5 is guidance and not legally binding. It states:
before an application, a (partial) QMS assessment by a notified body is not allowed, but "voluntary third-party activities carried out e.g. in the framework of certifications based on regulatory standards, including ISO 13485, are allowed" (Q I.6.2);
for suppliers, "the sole existence of a certificate against ISO 13485 is not sufficient" (Q IV.2).
EN ISO 13485, the European version of the standard, is covered in a separate article. Our article on the CE mark for a medical device covers the MDR route.
In our view, for EU devices that need a notified body, plan the ISO 13485 certificate and the MDR QMS assessment together, because they are separate activities.
MDSAP, a single audit accepted by several regulators, and the comparison with ISO 9001 are covered in separate articles.
A hypothetical example
Hypothetical example: a Dutch SaMD start-up plans to sell in the EU and the US.
Choice: it selects an accredited certification body and checks it in IAF CertSearch.
Initial audit: Stage 1 reviews its documentation and preparedness; Stage 2 assesses implementation at the sites to be on the certificate.
Cycle: it plans surveillance in years one and two and recertification in year three, using the MDSAP cycle as a reference.
US: it prepares for FDA inspection regardless of the certificate (FAQ Q13).
EU: it applies separately to a notified body for the MDR assessment (MDCG 2019-6, Q I.6.2).
Conclusion
ISO publishes the standard; accredited certification bodies certify. The certificate follows a three-year audit cycle, but it does not replace an FDA inspection or the notified body's MDR assessment.
In our view, choose the certification body for your markets, verify its accreditation, and treat the certificate as evidence, not as regulatory approval.
If you need help preparing for ISO 13485 certification, our management system team can help, or contact MedQAIR.
Frequently asked questions
Can ISO certify my company?
No. ISO does not perform certification or issue certificates. An independent certification body does.
Must the certification body be accredited?
Accreditation is not compulsory, according to ISO. Check whether the body is accredited through the national accreditation body or IAF CertSearch.
How do I verify a certificate?
Through IAF CertSearch or directly with the certification body.
Does FDA require ISO 13485 certification?
No. FDA will not require or issue ISO 13485 certificates, and a certificate does not exempt a manufacturer from FDA inspection (QMSR FAQ Q13).
Does a certificate replace the notified body audit?
No. Voluntary ISO 13485 certification is allowed before an application, but the notified body's MDR QMS assessment is a separate activity (MDCG 2019-6 Rev.5, Q I.6.2).
What happens in Stage 1?
Under the MDSAP approach, Stage 1 reviews the documentation and assesses preparedness for Stage 2.