Canada's Medical Devices Regulations (SOR/98-282) sort devices into Classes I to IV (section 6). Since 1 January 2019, MDSAP has replaced CMDCAS as the only quality system audit programme recognised by the Minister of Health, and Health Canada states that all licence holders now take part (Health Canada). On 1 April 2026, Health Canada published pre-market guidance for machine learning-enabled medical devices (MLMD), and amended establishment licence rules come into force on 14 December 2026.
The class decides how much Canada asks for. This article covers classification, who needs a medical device licence (MDL), what an application contains, the quality system certificate, changes, the establishment licence (MDEL), the 2026 fees and the MLMD guidance.
Classes and software
Under section 6, Class I is the lowest risk and Class IV the highest, using the rules in Schedule 1. Where more than one rule applies, the higher class applies (section 7).
Health Canada's SaMD guidance (effective 18 December 2019) explains how the rules apply to software:
| Rule | Condition | Class |
|---|---|---|
| Rule 10(1) | Software intended to image or monitor a physiological process | Class II |
| Rule 10(2) | As Rule 10(1), where erroneous readings could result in immediate danger | Class III |
| Rule 12 | Other software | Class I |
Source: Health Canada, Software as a Medical Device: Definition and Classification.
The class follows the intended use stated on the labelling. Software used in several situations takes the highest applicable class. Our articles on software as a medical device and medical device classifications compare the EU and US approaches.
Who needs a licence
Under section 26, no one may import or sell a Class II, III or IV device unless the manufacturer holds a licence for it, or an amended licence after a change under section 34. Class I devices do not need an MDL. Section 27 restricts advertising of Class II to IV devices that are not yet licensed.
In our view, keep two things apart: a licence covers the device, and an establishment licence covers the company that imports or sells it. Most foreign Class II to IV SaMD manufacturers only need the first (section 44(2)(c)).
What the application contains
| Item | Class II | Class III | Class IV |
|---|---|---|---|
| Device and manufacturer details (s. 32(1)(a)–(e)) | Yes | Yes | Yes |
| Intended use, list of standards, senior official attestation, label | Yes (s. 32(2)) | Yes | Yes |
| Quality system certificate | Yes | Yes | Yes |
| Foreign sales and recall history | No | Yes (s. 32(3)) | Yes |
| Summaries of studies and a bibliography | No | Yes | Yes |
| Risk assessment, quality plan and detailed studies, including software validation if appropriate | No | No | Yes (s. 32(4)) |
Source: SOR/98-282, section 32.
The quality system certificate
For Class II, III and IV, the application must include a CAN/CSA-ISO 13485 certificate issued by a recognised registrar (sections 32(2)(f), 32(3)(j), 32(4)(p) and 32.1). A certificate is valid for up to three years (section 32.2), and a new or modified certificate must be sent to Health Canada within 30 days (section 43.1).
In practice, the certificate comes from an MDSAP audit. Our article on MDSAP explains the programme, and our article on the ISO 13485 quality management system covers the standard.
In our view, the MDSAP certificate, not the application form, is the gating item for Class II to IV. Plan the audit first.
Changes and conditions
Under section 34, the manufacturer must apply for a licence amendment before making certain changes, including:
a significant change to a Class III or IV device;
a change that would affect the class;
a change to the manufacturer's name or the device name;
a change to the device identifier;
a change to the intended use of a Class II device.
Health Canada may also attach terms and conditions to a licence (section 36(2)).
The establishment licence (MDEL)
Under section 44, importers and sellers of medical devices generally need an MDEL. Exemptions include:
retailers;
health care facilities;
manufacturers of Class II to IV devices;
manufacturers of Class I devices that sell only through MDEL holders.
Under section 44(3), importers must carry out specific checks on the devices they bring in.
According to Health Canada's Phase 2 bulletin, from 14 December 2026:
foreign distributors that sell through a Canadian importer holding an MDEL no longer need their own MDEL;
requirements for documented procedures are clarified;
MDEL holders must keep a list of their suppliers.
Fees from 1 April 2026
| Fee | Amount (CAD) |
|---|---|
| MDL application, Class II | 643 |
| MDL application, Class III | 14,163 |
| MDL application, Class III near-patient IVD | 30,169 |
| MDL application, Class IV | 30,713 |
| Class II amendment | 331 |
| Private label | 179 |
| Annual right to sell | 460 |
| MDEL | 5,519 |
Sources: Health Canada, MDL application review fees, right to sell fees and MDEL fees.
Machine learning-enabled devices
Health Canada's pre-market guidance for MLMD (1 April 2026) applies to new and amendment applications for Class II to IV MLMD. It is guidance, not law. It describes a predetermined change control plan (PCCP) with three parts: a change description, a change protocol and an impact assessment. The cover letter should state that a PCCP is included.
Changes made under an authorised PCCP do not need a significant change amendment, but the other amendment rules still apply. Our guide to the predetermined change control plan covers the FDA version.
In our view, if the model will be retrained, build a PCCP into the first application.
A hypothetical example
Hypothetical example: a Dutch company makes machine learning software that tracks a physiological parameter.
Class: it classifies the software using Health Canada's SaMD guidance.
Certificate: it obtains its quality system certificate through an MDSAP audit.
Application: it applies under section 32, including a PCCP for retraining.
Sales: it sells through a Canadian importer that holds an MDEL.
Conclusion
In Canada, the class decides the licence and the content of the application, and Class II to IV need an MDSAP-based certificate. The device licence and the establishment licence are separate, and the MLMD guidance gives a route for planned model changes.
In our view, plan the MDSAP audit first, confirm who holds the MDEL in your supply chain, and include a PCCP if the model will change.
If you need help with a Health Canada licence, our global market registrations team can help, or contact MedQAIR.
Frequently asked questions
Does Class I software need a medical device licence?
No. Class I devices do not need an MDL, although the importer or seller may need an MDEL.
Does a foreign SaMD manufacturer need an MDEL?
Usually not. Manufacturers of Class II to IV devices are exempt from the MDEL requirement (section 44(2)(c)). Their Canadian importer or distributor may need one.
Is MDSAP mandatory?
For Class II to IV, the application needs a quality system certificate from a recognised registrar. MDSAP is the only programme recognised for this, so in practice, yes.
Does a model update need a licence amendment?
A significant change to a Class III or IV device does. Changes made under an authorised PCCP do not need a significant change amendment, but other amendment rules still apply.
How much does a Class II licence application cost?
CAD 643 from 1 April 2026, plus the annual right to sell fee of CAD 460.