ISO 14971:2019, the third edition, was published in December 2019 and confirmed in 2025. FDA recognised it in full on 23 December 2019 (recognition number 5-125). In the EU, Implementing Decision (EU) 2022/757 cited EN ISO 14971:2019 with A11:2021 under the MDR, with legal effect from 17 May 2022 (Commission summary list).
Reviewers judge the risk management process by its file. This article covers what the file must contain, how FDA and the MDR expect to see it, software and security risks, and how to keep the file current.
What ISO 14971 asks
According to FDA's recognition record and ISO, ISO 14971:2019:
requires the manufacturer to identify hazards, estimate and evaluate the risks, control them and monitor the effectiveness of the controls;
applies to all stages of the device life cycle;
does not set acceptable risk levels; the manufacturer sets objective criteria for risk acceptability;
does not require a quality management system;
includes risks related to data and systems security and to usability.
In our view, the acceptability criteria drive every later decision, so fix them before the first risk evaluation.
What the file contains
| Element | What it holds | Source |
|---|---|---|
| Risk management plan | The criteria for risk acceptability and the method for evaluating overall residual risk against the benefits, documented before the first evaluation | FDA software guidance, VI.C |
| Risk assessment | Hazards, including reasonably foreseeable misuse; sequences of events; hazardous situations; severity and risk estimation; initial risk evaluation | FDA software guidance, VI.C |
| Risk controls | Controls in priority order: inherent safety by design, protective measures, then information for safety | FDA software guidance, VI.C |
| Verification of controls | Evidence that each control is implemented and effective, traced to requirements and tests; any new hazards introduced by the controls | FDA software guidance, VI.C |
| Residual risk | Evaluation of residual risk, and a benefit-risk analysis where residual risk is not acceptable | FDA software guidance, VI.C |
| Risk management report | Confirmation that the plan was implemented, that the file was assessed by appropriate personnel, that overall residual risk is acceptable, and that methods are in place to collect production and post-production information | FDA software guidance, VI.C |
Sources: FDA, Content of Premarket Submissions for Device Software Functions (June 2023), Section VI.C; ISO 14971:2019.
In our view, treat the file as an index to controlled records, not as one long document. Our article on verification and validation covers how control verification is evidenced, and our article on SaMD technical documentation shows where the file sits.
Software failure probability
FDA notes that the probability of a software failure can be hard to estimate, and that setting it to 1 can be appropriate (software guidance, Section VI.C, footnote 45). The sequence of events from the failure to the hazardous situation and harm still needs to be documented.
The US view
The QMSR incorporates ISO 13485:2016, which FDA says "now specifically requires risk management". FDA's draft guidance on quality management system information for PMA and HDE submissions (October 2025, not for implementation) describes Subclause 7.1 as requiring risk management processes in product realisation, with records kept.
FDA recognises ISO 14971:2019 completely; ANSI/AAMI ISO 14971:2019 is the identical US adoption. Our article on the ISO 13485 quality management system covers the wider system.
The EU view
Under the MDR, Annex I Section 3 requires a risk management system that runs through the whole life cycle as a continuous iterative process, and Section 4 covers risk control measures. Sections 1, 2 and 8 cover performance, reducing risks as far as possible and minimising foreseeable risks (MDCG 2019-16 Rev.1, Table 1). The technical documentation must include the justification, validation and verification of the solutions adopted.
Harmonised standards are voluntary; once cited in the Official Journal, they give a presumption of conformity with the requirements they cover. Our article on the EU MDR for SaMD explains the wider framework.
ISO/TR 24971
ISO/TR 24971:2020 gives informative guidance on applying ISO 14971. The second edition was published in June 2020, and ISO lists it as "to be revised". Annex E covers in vitro diagnostic devices. FDA lists it as a supportive document.
Security risks
According to MDCG 2019-16 (Section 3.2), security risk management has the same process elements as safety risk management, and security risks that can affect safety must be considered in the safety risk assessment, and the other way round. FDA's recognition record notes that its cybersecurity guidance (February 2026) uses exploitability rather than probability for security risks.
In our view, keep safety and security risk records linked in both directions.
Keeping the file current
The risk management report confirms that methods are in place to collect production and post-production information. That information feeds back into the file, so the file is updated throughout the life of the device. Our article on post-market surveillance covers where that information comes from.
A hypothetical example
Hypothetical example: a Dutch company makes cloud-based SaMD.
Plan: it sets the acceptability criteria before the first evaluation.
Hazards: it identifies hazards, including reasonably foreseeable misuse.
Software: where software failure probability cannot be estimated, it sets it to 1 and documents the event sequence.
Controls: it applies controls in FDA's priority order and traces each one to the software requirements, design and tests.
Security: it cross-references security risks with the safety assessment.
Report: it confirms that overall residual risk is acceptable.
After launch: post-market data updates the file.
Conclusion
The risk management file shows that the ISO 14971 process was followed: criteria first, then hazards, controls, verification, residual risk and a report. FDA and the MDR both expect it, and it has to stay current after launch.
In our view, build the file as an index of controlled records, fix the acceptability criteria early, and link safety and security risks.
If you need help with your risk management file, our management system team can help, or contact MedQAIR.
Frequently asked questions
Is the risk management file one document?
No. It can be a set of records, or an index pointing to them, as long as it shows the whole process.
Does FDA want risk management information in a premarket submission?
For device software, FDA's June 2023 software guidance describes risk management documentation as part of the submission, including the plan, assessment and report.
Is EN ISO 14971 mandatory under the MDR?
No. Harmonised standards are voluntary. Using EN ISO 14971:2019 with A11:2021 gives a presumption of conformity with the requirements it covers.
What does ISO/TR 24971 add?
Informative guidance on applying ISO 14971. It is not a requirement.
Can software failure probability be set to 1?
Yes. FDA notes that setting it to 1 can be appropriate where it cannot be estimated.