Verification and Validation of Medical Devices Explained

Software engineer running verification tests and clinician validating the same medical device software in clinical use

The QMSR took effect on 2 February 2026 and incorporates ISO 13485:2016 by reference. FDA's premarket software guidance now notes that the QMSR does not use the terms "Design Controls" and "Design Validation"; their elements sit in ISO 13485 Clause 7.3 and its subclauses.

When verification and validation (V&V) evidence is blurred, it shows up in submissions and audits. This article covers the definitions, who must comply, ISO 13485 Clauses 7.3.6 and 7.3.7, software V&V for FDA, and how the EU MDR treats V&V.

Definitions

Under the QMSR, 21 CFR 820.3 applies the definitions in ISO 13485 and ISO 9000:2015 Clause 3. FDA's software guidance (Section IV) defines the terms for software:

  • Verification: "confirmation by objective evidence that the output of a particular phase of development meets all the input requirements for that phase".

  • Validation: "establishing, by objective evidence, that the software specifications conform to user needs and intended uses, and that the particular requirements implemented through software can be consistently fulfilled".

The distinction is not new: the former Quality System regulation was harmonised with ISO 8402:1994, which treated the two terms as distinct (General Principles of Software Validation, Section 3.1.2).

In our view, verification shows the outputs meet the inputs, and validation shows the device meets user needs and its intended use. One test can serve both only if the traceability shows it.

Who must comply

Under 21 CFR 820.10(c), ISO 13485 Clause 7.3 applies to class II and class III devices, class I devices automated with computer software, and the listed class I devices. According to FDA's QMSR page, devices under an IDE are not exempt from these requirements. Our article on 21 CFR Part 820 explains the QMSR, and our article on the ISO 13485 quality management system covers the standard.

Clauses 7.3.6 and 7.3.7

FDA's draft guidance on quality management system information for PMA and HDE submissions (October 2025) describes what FDA expects for the two clauses. It is a draft and not for implementation.

Aspect Verification (7.3.6) Validation (7.3.7) Source
Question answered Do the design outputs meet the design inputs? Does the device meet user needs and its intended use? FDA software guidance
Compared against Input requirements User needs and intended uses FDA software guidance
What FDA lists Verification activities and reports; methods, protocols and acceptance criteria; statistical techniques with a rationale for sample size Validation activities and reports; evidence that results meet the intended use; plans with methods, acceptance criteria and statistical techniques FDA draft QMS information guidance
Software example Code review; unit, integration and system testing Testing with intended users in actual or simulated use FDA software guidance

Sources: FDA draft QMS information guidance (October 2025); FDA, Content of Premarket Submissions for Device Software Functions (June 2023).

Software V&V for FDA

FDA's June 2023 software guidance sets two documentation levels. For Basic documentation, the V&V content is a summary of unit, integration and system testing with the software version and pass/fail results, fixes after failed tests, a regression analysis, the system-level test protocol and report, and unresolved anomalies deferred on the basis of a risk assessment. Enhanced documentation adds the unit and integration test protocols and reports. Both include the software version history.

The General Principles of Software Validation (January 2002) treat software validation as part of design validation. Its Section 6 has been superseded by the Computer Software Assurance guidance, which covers production and quality system software. Our article on FDA software validation guidance explains which document applies to which software.

Risk controls need both: verification that each control is implemented and evidence that it is effective, traced from the hazard to the software requirements and design and to the tests (software guidance, Section VI.C).

In our view, derive software V&V from the requirements and the risk file. The traceability is the evidence.

The EU MDR view

According to MDCG 2019-16 Rev.1 (Table 1), MDR Annex I Section 17.2 requires software to be developed according to the state of the art, taking into account the development life cycle, risk management, information security, verification and validation. The technical documentation must include the justification, validation and verification of the solutions adopted (Annex II Section 4). Our article on SaMD technical documentation shows where these records sit in the technical file.

MDCG 2020-1 describes technical performance evidence from V&V activities such as unit, integration and system testing (Section 4.3). For clinical performance validation, it says this should be considered at each new release; if it is not repeated, a justification should be recorded in the technical documentation (Section 4.4).

Clinical evaluation is a separate MDR process from design validation (Article 2(44), as quoted in MDCG 2020-1). Our article on the clinical evaluation report covers it. Usability is covered in our article on FDA human factors guidance.

In our view, one V&V record set can serve both FDA and the MDR, built on one traceability chain with a summary for each.

A hypothetical example

Hypothetical example: a Dutch company develops image-analysis SaMD.

  1. Traceability: it traces each requirement to the design, the tests and the risk controls.

  2. Verification: it carries out code review and unit, integration and system testing.

  3. Validation: it validates the software in simulated use with intended users.

  4. Risk controls: it verifies each control and traces it to its hazard.

  5. FDA: it determines its documentation level under the software guidance and documents the reasoning.

  6. EU: it presents technical and clinical performance evidence as described in MDCG 2020-1.

  7. New release: it runs a regression analysis and records its decision on clinical performance validation.

Conclusion

Verification proves the outputs meet the inputs; validation proves the device meets user needs and its intended use. Under the QMSR, both sit in ISO 13485 Clause 7.3, and the MDR asks for the same evidence in the technical documentation.

In our view, keep one traceability chain from user needs to tests and risk controls, and decide at each release what must be repeated.

If you need help planning V&V for your software, our software lifecycle management team can help, or contact MedQAIR.

Frequently asked questions

Is software testing verification or validation?

Mostly verification: unit, integration and system testing show the software meets its requirements. Validation shows the software meets user needs and its intended use, for example through testing with intended users.

Is clinical evaluation the same as design validation?

No. Under the MDR, clinical evaluation is a separate process (Article 2(44)), although it can use evidence from validation.

Does V&V have to be repeated for each release?

The affected V&V should be repeated, with a regression analysis. MDCG 2020-1 says clinical performance validation should be considered at each new release, with a justification if it is not repeated.

Is FDA's October 2025 QMS information guidance binding?

No. It is a draft and not for implementation. The binding requirements are in ISO 13485 Clause 7.3 as incorporated into 21 CFR Part 820.

Latest Blogs

November 23, 2026

The QMSR took effect on 2 February 2026 and incorporates ISO 13485:2016 by reference. FDA's premarket software guidance now notes

November 22, 2026

On 29 May 2025, FDA issued the final guidance Requests for Feedback and Meetings for Medical Device Submissions: The Q-Submission

November 21, 2026

The MDR of 5 April 2017 requires every manufacturer's quality management system to include "a strategy for regulatory compliance, including

logo

Unlock Your Quick Guide to AI Act
Compliance!

Explore AI-enabled SaMD requirements with our easy step-by-step guide.

Cookies help us improve your experience on our website. By using our site, you consent to the use of cookies as described in this policy.